Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

Date

Issue

Symptoms

Corrective Action

Additional Information

4/19/22

Mount did not go active

Aborted-Error 2

az failed with error(-14) controller interlocked

alt failed with error(-14) controller interlocked

Beckhoff error: external error

Reset Phase guard in MDC2 (service ring) cycle 3p circuit breaker above unit.

Phase guard was replaced 7/26/21 continue to monitor.

4/19/22

M3 fault

aborted- Hardware exceeption when setting PI E-712 servo mode

Reset PI controller in PA&C rack.

IT working on ports may have caused issue.

4/19/22

Pflow Coude door stuck open

Pflow Coude lab Roll up door doesn’t close (stuck open)

Re-stripped switch wires and terminate

Switch wires oxidized