This page is to help identify issues with the operation on the telescope and corrective action.
...
Date
...
Issue
...
Symptoms
...
Corrective Action
...
Filter by label (Content by label) | ||||||||||
---|---|---|---|---|---|---|---|---|---|---|
|
1 | 4/19/22 | Mount did not go active | Aborted-Error 2 |
Az failed with error(-14) controller interlocked |
Alt failed with error(-14) controller interlocked Beckhoff error: external error | Reset Phase guard in MDC2 (service ring) cycle 3p circuit breaker above unit. | Phase guard was replaced 7/26/21 continue to monitor. | |||
2 | 4/19/22 | M3 fault | aborted- Hardware exceeption when setting PI E-712 servo mode | Reset PI controller in PA&C rack. | IT working on ports may have caused issue. |
3 | 4/19/22 | Pflow Coude door stuck open | Pflow Coude lab Roll up door doesn’t close (stuck open) | Re-stripped switch wires and terminate | Switch wires oxidized |
4 | 10/17/22 | Mount Motor Drive Fault. | MDC2 Phase Guard fault. Reset and it cleared. |